|  |  |  | APPENDIX 2 - DXA quality control
 
        At time of installation, (as part of commissioning procedure) or 
          after any major maintenance procedure or major software upgrade 
          
            Machine calibration and testing by supplier. Accuracy 
              and precision evaluationIn vitro: short-term precisionScan manufacturer's recommended QC phantom 10-20 times. From these 
              measurements calculate the mean BMD of the Phantom and the coefficient 
              of variation (CV) of the results. (See Appendix 
              3).
 CV = Standard deviation x 100%/Mean
 
In vivo: short-term precisionOn each machine, obtain voluntary followup scans within a short 
              timeframe (maximum 1 month), or duplicate scans on the same day, 
              on at least 28 subjects. The volunteers should be in an age range 
              appropriate to the generally scanned population and should give 
              informed consent. From these values obtain the in-vivo short-term 
              precision for BMD (See Appendix 4).
 
 
Ongoing quality control proceduresCalibration and quality control according to manufacturer's specifications. 
          The QC phantom shall be scanned at least twice weekly (and preferably 
          daily) using the same scanning parameters. This phantom is not the daily 
          calibration phantom, but is an anthropomorphic (or quasi-anthropomorphic) 
          phantom recommended by (or at least acceptable to) the manufacturer. 
          The QC phantom data shall be recorded and checked for medium-term precisional 
          error and systematic bias using the appropriate statistical analysis. 
          This analysis may include one or more of the following tests: Multi-rule 
          Shewart Chart analysis (Appendix 5), CUSUM (Appendix 
          6), and Running Mean. Further information to assist in the choice 
          of statistical methods (particularly regarding reporting of results) 
          is given in Appendix 3. The ongoing QC program 
          must include a regular maintenance schedule, with periodic testing of 
          accuracy using a suitable QC phantom, as recommended by the manufacturer. 
          This phantom may be the same as that used for the regular (at least 
          twice weekly) QC. All interventions by the maintenance/repair provider 
          shall be recorded so that variations in QC parameters may be correlated 
          with breakdowns and repairs if necessary - partly to facilitate retrospective 
          analysis and correction of BMD data, if this should, with hindsight, 
          be deemed necessary.
 
 
Long term Quality ControlExamination of trends in QC parameters (particularly BMD) over periods 
          of years is a necessary precaution for establishing the integrity of 
          long term studies of individual patients, or for the reporting of results 
          from clinical trials. The statistical methods used to examine long term 
          trends are the same as those use for medium term studies (see above). 
          However, it should be noted that statistical analyses become more powerful 
          and illuminating, with the acquisition of larger data groups - so that 
          analyses are worth periodically repeating as the QC history of a machine 
          accumulates.
 
 Each machine has specific operating characteristics which leads to some 
          QC parameters being more important than others, depending on the circumstances. 
          For example, experience with Hologic pencil beam systems has shown that 
          variations in the parameters "k" and "d0" are frequently 
          more sensitivity to the onset of malfunctions than are BMD, BMC or Area. 
          Such information should be incorporated into the operations manual for 
          a machine.
 
 
Use of internationally accepted anthropomorphic 
          standards in ongoing multi-centre QC studiesIt is likely that some form of multi-centre participatory QC will eventually 
          be either strongly recommended or mandatory. The use of such standards 
          allows the BMD of a particular centre to be expressed as a machine- 
          and location-independent "standardised" BMD (sBMD), which 
          is much more amenable to allowing comparisons between centres in multi-centre 
          clinical trials.
 
 However, sBMD is not currently recommended to compare the measurements 
          of the same patients between machines or between centres.
 
 
 |  |